top of page

Sysmex CN-3000/6000

Sysmex Corporation. Japan
image.png
DESIGNED FOR VERSATILITY AND EXCELLENCE IN THE LABORATORY

The smallest footprint, highest throughput, and most comprehensive test menu on a single analyzer—ranging from routine to specialized tests (e.g., DOAC and platelet function assays/aggregatometry).

The minimum plasma volume required for analysis has been reduced by 65% ​​through the implementation of a new, improved sample dosing system.

Автономні прилади, що окремо стоять, у стандартній комплектації

Enables the use of various analytical methods—such as the analysis of inhibitory activity via cross-mixing tests and Reaction Curve Analysis (CWA) for detecting abnormal clot formation kinetics—which may be observed in patients with coagulopathies.

Automated coagulometer with capabilities for preliminary sample quality assessment and platelet function analysis

Measuring principles: clotting, chromogenic, immunochemical and aggregation

38 positions for reagents (with cooling capability to 4–12°C) and 6 positions for buffers (room temperature)

A modular system that scales with your needs

The CN-3000 and CN-6000 can be flexibly configured to meet the individual needs of laboratory automation.

Test volume forecasting function; automated dilution of platelet aggregation agonists

Combines routine and specialized tests—including aggregometry tests—in a single analyzer.

Control of pre-analytical and analytical phases

Automated control of the pre-analytical phase: preliminary screening of samples for hemolysis, icterus, and chyle (HIL detector), as well as assessment of the correct anticoagulant-to-blood ratio (sample volume detector).

Multi-wavelength technology that automatically adapts the analyzer’s optical system based on sample properties (analyte concentration, presence of hemolysis, turbidity, or icterus), ensuring highly accurate results even when processing complex samples.

Process automation

Capability to work with primary tubes; samples can be placed into the analyzer immediately after centrifugation, without requiring additional handling.

The cap-piercing function allows for the use of primary tubes and eliminates direct contact with biological material.

Full automation of the sample dilution system, both for determining factor or inhibitory activity in plasma and for constructing calibration curves.

Автоматична система попереднього розведення реагентів, виходячи з прогнозованої оцінки кількості тестів

Reagents, samples, and cuvettes are loaded during operation, ensuring continuous testing regardless of the workload.

CHARACTERISTICS

Measurement principle

Reagents

Parameters under investigation

​Wavelengths

Productivity

Measuring channels

Number of samples

Calibration

Memory

Cuvettes

Quality Control

Dimensions

The service life of the lid-piercing system has been increased fourfold compared to previous models.

The optical system uses an LED, which has increased the light source's lifespan to 36,000 hours.

Reagent vials are positioned at an angle in the rack, which reduces dead volume and increases the number of tests that can be performed with each vial.

image.png

Measurement principles: clotting, chromogenic, immunochemical, and aggregation

38 positions for reagents (with cooling capability to 4–12°C) and 6 positions for buffers (room temperature)

Clotting assays: PT, aPTT, Fibrinogen, Batroxobin (Reptilase) time, TT, factors of the extrinsic and intrinsic pathways (II, V, VII, X, VIII, IX, XI, XII), Protein S, Protein C, Factor V Leiden, Lupus Anticoagulant (LA), ProC Global, Rivaroxaban, Apixaban, Edoxaban, Dabigatran

Chromogenic assays: Antithrombin III (AT-III), Plasminogen (PLG), Alpha-2-antiplasmin (α2-AP), Protein C (PC), FVIII, FXIII, PAI levels and activity, Heparin, C1-inhibitor, etc. Immunological assays: D-dimer, von Willebrand factor concentration and activity (vWF:Ag, vWF:Ac), etc.

Aggregometry: assessment of platelet function using various inducers (ADP, arachidonic acid, collagen, ristocetin), evaluation of the efficacy of antiplatelet therapy (aspirin, clopidogrel), and ristocetin-based von Willebrand factor activity.

Multi-wavelength system for simultaneous sample scanning at 5 wavelengths (340, 405, 575, 660, and 800 nm)

CN-3000:

PT: 225 tests/hour, PT+APTT: 215 tests/hour

CN-6000:

PT: 450 tests/hour

PT+APTT: tests/hour

CN-3000 – 13 analytical channels (4 with aggregometry capability)

CN-6000 – 26 analytical channels (8 with aggregometry capability)

Simultaneous loading of 30 samples as standard; capacity expandable to 120 samples using the SA-90 auto-loader system (optional)

Cap-piercing capability; positions for emergency (STAT) samples

Simultaneous availability of calibration curves (2–12 points) for 250 methods

Up to 10,000 samples

Simultaneous loading of up to 1,200 cuvettes, with the capability for continuous reloading.

Automated quality control program. X-bar control, Levey-Jennings, and Westgard rules monitoring. Up to 40 control files, with 1,200 data points for each.

595 x 1350 x 906 cm, 230 kg (PU 17 compressor included; computer and monitor not included).