Evidence MultiSTAT
Автоматизоване тестування на наркотики
System Features
Biological fluids: blood, urine, saliva
Time to result: 17–30 minutes
Biochip Array Technology
Simultaneous detection of up to 29 classic, prescription, and synthetic drugs
Sample volume: 170–200 µL
System Features
Utilizing revolutionary Biochip Array Technology (BAT), the Evidence MultiSTAT is an automated analyzer that enables the simultaneous, on-site detection of up to 29 classic, prescription, and synthetic drugs from a single sample. Designed to process a wide variety of sample matrices, our patented multi-analyte testing platform delivers a comprehensive toxicological profile in a matter of minutes—fundamentally transforming the landscape of drug detection.
.png)
7x7 – biochip surface
44 separate test zones on each biochip
BAT is an immunoassay testing platform for the simultaneous detection of multiple analytes within a panel of related tests. The technology works by combining a panel of up to 44 related tests on a single biochip, using a single set of reagents, controls, and calibrators.
A competitive chemiluminescent immunoassay method is used for biochip arrays. The light signal generated from each of the test areas on the biochip is simultaneously detected using digital imaging technology and compared against a calibration curve signal or a cut-off value.
Cartridge
.png)
1
The first well, into which the cut-off, control, calibrator, or test sample is added, depending on the kit.
2
The second well, into which a control, calibrator, or sample is added.
3
The area containing all additional liquids (solvents, reagents) is covered with foil.
4
The arrangement of two biochips, each containing 44 separate test zones.
Analyzer
1
A large touchscreen allows the user to easily switch between the test screen and the view for sample and control results.
2
Position for the user to load pre-filled cartridges with tips.
3
Positioning of the cartridge loading mechanism for sample cartridges prior to testing.
4
Two additional USB ports for connecting peripheral equipment
.png)
Dynamics of the study process within the analyzer
.png)
The solvent, sample, and conjugate are added to the biochip well.
The biochip is being incubated.
Biochip flushing
Results available in 17–30 minutes
The biochip is loaded into the fixation zone.
A signaling reagent is added to the biochip well.
.png)