Evidence MultiSTAT

Автоматизоване тестування на наркотики

System Features

Biological fluids: blood, urine, saliva

Time to result: 17–30 minutes

Biochip Array Technology

Simultaneous detection of up to 29 classic, prescription, and synthetic drugs

Sample volume: 170–200 µL

System Features

Utilizing revolutionary Biochip Array Technology (BAT), the Evidence MultiSTAT is an automated analyzer that enables the simultaneous, on-site detection of up to 29 classic, prescription, and synthetic drugs from a single sample. Designed to process a wide variety of sample matrices, our patented multi-analyte testing platform delivers a comprehensive toxicological profile in a matter of minutes—fundamentally transforming the landscape of drug detection.

1 (4).png

7x7 – biochip surface

44 separate test zones on each biochip

BAT is an immunoassay testing platform for the simultaneous detection of multiple analytes within a panel of related tests. The technology works by combining a panel of up to 44 related tests on a single biochip, using a single set of reagents, controls, and calibrators.

A competitive chemiluminescent immunoassay method is used for biochip arrays. The light signal generated from each of the test areas on the biochip is simultaneously detected using digital imaging technology and compared against a calibration curve signal or a cut-off value.

Cartridge
1 (5).png

1

The first well, into which the cut-off, control, calibrator, or test sample is added, depending on the kit.

2

The second well, into which a control, calibrator, or sample is added.

3

The area containing all additional liquids (solvents, reagents) is covered with foil.

4

The arrangement of two biochips, each containing 44 separate test zones.

Analyzer

1

A large touchscreen allows the user to easily switch between the test screen and the view for sample and control results.

2

Position for the user to load pre-filled cartridges with tips.

3

Positioning of the cartridge loading mechanism for sample cartridges prior to testing.

4

Two additional USB ports for connecting peripheral equipment

1 (6).png
Dynamics of the study process within the analyzer
1 (7).png

The solvent, sample, and conjugate are added to the biochip well.

The biochip is being incubated.

Biochip flushing

Results available in 17–30 minutes

The biochip is loaded into the fixation zone.

A signaling reagent is added to the biochip well.

1 (8).png