top of page

Sysmex CS-2400/2500

Sysmex Corporation. Japan
image.png
DESIGNED FOR VERSATILITY AND EXCELLENCE IN THE LABORATORY

The CS-2500 analyzer is capable of simultaneously performing tests to evaluate all components of hemostasis—including primary vascular-platelet hemostasis, secondary hemostasis, and fibrinolysis—as well as immunoturbidimetric tests.

Multi-wavelength technology, featuring an optical system that automatically adapts based on sample properties, enables highly accurate results—even when working with "difficult" samples.

The optical system comprises 10 independent analytical channels, allowing the CS-2500 analyzer to perform tests using all supported methodologies simultaneously at no fewer than five wavelengths.

Automated high-precision testing of platelet function using aggregometry

Improved quality of results for pathological samples; automated retesting and reflex testing rules integrated into the software.

Process automation

Thanks to the capability to process primary tubes, samples can be placed into the analyzer immediately after centrifugation, without any additional handling.

Reagents, samples, and cuvettes are loaded during operation, ensuring continuous testing regardless of the workload.

The capacity to load 500 cuvettes on board ensures long-term instrument operation without operator intervention.

When upper and lower limits for re-analysis are set, the system automatically performs a re-analysis—using new dilutions, if necessary.

Automated hemostasis system analyzer with aggregometer function

Combining both routine hemostasis assessment tests and highly specialized assays within a single analytical system

Automatic sample quality control

Improved quality of results for pathological samples

Automated high-precision platelet function testing

Extended assessments of hemostasis inhibitors

Throughput of up to 180 tests per hour for clotting assays, with the capability to analyze a sample for 60 parameters.

Control of pre-analytical and analytical phases

Determination of the correctness of the blood-to-anticoagulant ratio, with the option (at the user's discretion) to accept or reject the sample for further analysis.

The optical system of the CS-2500 analyzer allows for the assessment of the level of hemolysis, icterus, and lipemia in a sample (HIL detector), with subsequent flagging of results that do not meet user-defined quality criteria.

One of the sections of each reagent rack allows for the mixing of reagents.

The cap-piercing function eliminates direct contact with biological material, significantly reducing the risk of infection via airborne transmission.

Reagent racks are placed on a dual, rotating, thermostated tray with 40 positions.

The vials are slightly tilted to reduce dead volume and increase the number of tests per vial, thereby enhancing the economic efficiency of operating the analyzer.

CHARACTERISTICS

Measurement principle

Clotting method

Chromogenic substrate method

Immunoturbidimetry

Reagents

Parameters under investigation

​Wavelengths

40 positions, with thermostat or cooling function, as well as stirring

Over 60 different tests: PT, aPTT, fibrinogen, extrinsic pathway factors (II, V, VII, X), intrinsic pathway factors (VIII, IX, XI, XII), factor XIII (chromogenic), AT-III, α2-antiplasmin, plasminogen, protein C, protein S, fibrin degradation products, von Willebrand factor, D-dimer, platelet aggregation analysis using various inducers, etc.

Multi-wavelength system for simultaneous sample scanning at 5 wavelengths: 340, 405, 575, 660, and 800 nm

Productivity

PT: up to 180 tests per hour

PT + APTT: up to 180 tests per hour

PT + APTT + optical fibrinogen: up to 180 tests per hour

PT + APTT + optical fibrinogen + antithrombin + D-dimer: up to 95 tests per hour

Measuring channels

10 analytical channels. 4 of these feature a stirring function for aggregometry, enabling the simultaneous use of a full range of analytical methods: clotting, chromogenic, immunoturbidimetric, and aggregation assays.

Number of samples

50 samples on board simultaneously, in primary or secondary tubes

5 separate positions for urgent samples

Sample size

From 5 µL. Aspiration of samples and reagents using a sampler with automatic liquid level detection in the tube.

Cuvettes

Quality Control

Dimensions

Simultaneous loading of 500 cuvettes with automatic feed

Levey-Jennings 1200 points × 750 files, 40 parameters

main block: 775 x 675 x 865 / 100 kg

pneumatic block: 280 x 400 x 355 mm / 17 kg