top of page

Sysmex XN-3100 Hematology Station

Sysmex Corporation. Japan
image.png
DESIGNED FOR VERSATILITY

The XN-3100 hematology system ensures standardization of the entire complete blood count (CBC) process, including the preparation and staining of blood smears.

The station comprises two automated hematology analysis modules—each of which can be equipped with the full range of diagnostic capabilities of the XN series—and an SP-50 slide preparation and staining module (for the XN-3100 station).

The inclusion of two hematology analysis modules in the station allows for the functional separation of tasks between them: a screening module and an advanced hematology analysis module.

The modular division of functions, combined with the effective application of reflex testing, allows the analysis algorithm to be adapted individually for each sample.

By employing a system of rules, the station enables functional rationalization and, consequently, a significant reduction in the volume of microscopy performed.

The SP-50 slide staining module can be used independently for the preparation of peripheral blood smears, as well as for staining bone marrow smears.

Overall control of the station via a computer-based station (extended IPU) ensures the management of all processes.

High-throughput modular hematology station with smear preparation and staining functionality

The system consists of two hematology modules and a smear preparation and staining station.

The station’s two-module design implements the concept of an integrated back-up.

The throughput of the smear preparation and staining module can be adapted to the laboratory's needs.

Full automation of the hematological analysis process, including smear preparation and staining, with the implementation of reflex testing.

image.png
image.png
Configurations

XN-10 automated hematology analysis module:

1. Basic CBC+DIFF mode

2. Optional diagnostic channels:

RET channel (analysis of reticulocytes and reticulocyte parameters)

PLT-F channel (fluorescent platelet analysis with assessment of the immature platelet fraction)

BF channel (cytological analysis of body fluids)

XN-20 Automated Hematology Analysis Module:

1. Basic mode: CBC+DIFF, RET channel (analysis of reticulocytes and reticulocyte parameters), WPC channel (channel for diagnosing neoplastic processes and counting hematopoietic precursor cells)

2. Optional diagnostic channels:

PLT-F channel (fluorescent platelet analysis with assessment of the immature platelet fraction)

BF channel (cytological analysis of body fluids)

Diagnostic modules

Модуль гематологічного аналізу XN-10/20

CBC+DIFF – режим повного гематологічного аналізу крові з диференціюванням лейкоцитарної формули на 6 субпопуляцій лейкоцитів: 200 зразків/год

CBC+DIFF+RET – режим повного гематологічного аналізу крові з диференціюванням лейкоцитарної формули на 6 субпопуляцій лейкоцитів і аналіз ретикулоцитів (активовані 2 RET каналу): 163 зразка/год

​BF-аналіз – цитологічне дослідження рідин тіла (активовані 2 BF-каналу): 80 зразків/год

SP-50 automated smear preparation and staining module (XN-3100):

standard mode — 35 stained smears per hour

high-throughput mode (high throughput licence) — 75 stained smears per hour

SP-50 Automatic Smear Preparation and Staining Module

DI-60 digital imaging and morphological analysis module for blood cells and body fluids – optional

CHARACTERISTICS

Measurement principle

CBC+DIFF mode parameters

RET channel parameters

​PLT-F-channel parameters

​WPC channel parameters

​BF channel parameters

Throughput capacity

Sample aspiration volume

Device memory

Quality Control

Reagents

Dimensions

WBC DIFF/RET/PLT-F/WPC: Flow cytometry

RBC/PLT: Hydrodynamic focusing with the impedance method

HGB (hemoglobin): Cyanide-free SLS method

Smear preparation: Wedge method

Smear staining protocols: Romanowsky-Giemsa, Wright-Giemsa, and Wright methods

Image analysis system (optional): Digital microscopy method with image analysis using artificial neural networks

​WBC, NRBC%, NRBC#, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, RDW-SD, RDW-CV, MicroR%, MacroR%, PDW, MPV, P-LCR, PCT, NEUT%, NEUT#, LYMPH%, LYMPH#, MONO%, MONO#, EO%, EO#, BASO%, IG#

Optional: Extended Inflammation Parameters license: NEUT-GI, NEUT-RI, RE-LYMP%, RE-LYMP#, AS-LYMP%, AS LYMP#

Optional: TNC, TNC-C, TNC-D, WBCC, WBC-D, NE-SSC, NE-SFL, NEFSC, HFLC#, HFLC%

​RET%, RET#, IRF, LFR, MFR, HFR, RET-He, RBC-He, Delta-He, HYPO-He%, HYPER He%, PLT-O

Additional: FRC#, FRC%, RPI, RBC-O, HGB O, MCHC-O

PLT-F, IPF, IPF#

Highly specific marker for neoplastic processes

Optional: stem cell count (XN Stem Cells licence): HPC%, HPC#

​Main: WBC-BF, RBC-BF, MN%, MN#, PMN%, PMN#, TC-BF#

Additional: HF-BF#, HF-BF%, NE-BF#, NE BF%, LY-BF#, LY-BF%, MOBF#, MO-BF%, EO-BF#, EO-BF%, RBC-BF-2

CBC+DIFF – complete blood count mode with a 6-part white blood cell differential: 200 samples/hour

CBC+DIFF+RET – complete blood count mode with a 6-part white blood cell differential and reticulocyte analysis (2 RET channels activated): 163 samples/hour

BF analysis – cytological examination of body fluids (2 BF channels activated): 80 samples/hour

SP-50 automated slide preparation and staining module (XN-3100):

standard mode — 35 stained slides per hour

high-throughput mode (high throughput license) — 75 stained slides per hour

Модуль гематологічного аналізу (XN-10/20):

88 мкл венозної або капілярна кров з ЕДТА для режимів CBC+DIFF, CBC+DIFF+RET і CBC+DIFF+RET+PLT-F

​88 мкл зразка біологічної рідини для режиму BF-аналіз

Модуль приготування та фарбування мазка SP-50 (для XN-3100):

70 мкл цільної крові з ЕДТА для автоматичної подачі

​35 мкл крові з ЕДТА для мануального режиму подачі зразка

Volume of stored information: complete test results for 100,000 samples

Patient information: 100,000 records

Quality control (QC) data: 99 files (300 charts per file); X-bar control charts (Levey-Jennings charts) — 94 files (300 charts per file); X-barM control charts — 5 files (300 charts per file)

History of reagent replacements: 5,000 records

XN CHECK L1, XN CHECK L2, XN CHECK L3: Material (3 levels) for quality control of all parameters measured in CBC+DIFF+RET mode. Pack size: 3 ml per level.

XN Check BF: Material (2 levels) for quality control of all parameters measured in CBC+DIFF+RET mode. Pack size: 3 ml per level.

SNCS — an external international quality control system available to all users of Sysmex analyzers.

​SVS+DIFF mode, BF mode:

CELLPACK DCL 10/20 l

SULFOLYSER 3 x 500 ml / 5 l

LYSERCELL WDF 2 l

FLUOROCELL WDF 2 x 22 ml

LYSERCELL WNR 5L

FLUOROCELL WNR 2X82ML

CELLCLEAN 50 ml

RET mode:

CELLPACK DFL 2 x 1.5 l

FLUOROCELL RET 2 x 12 ml

PLT-F mode:

CELLPACK DFL 2 x 1.5 l

Fluorocell PLT 2 x 12 ml

WPC mode:

Lysercell WPC 2 x 1.5 l

Fluorocell WPC 2 x 12 ml

XN-3100: 2,000 x 921 x 1,083 mm / 320 kg, including 3 modules (2 x XN-10/20 and SP 50) and a shared automated sample loader

bottom of page