Sysmex XN 330
Sysmex Corporation. Japan

Automated 6-diff hematology analyzer
A budget-friendly solution for top-tier hematological analysis
Best-in-class solution for small and medium-sized laboratories
Complete blood count from 25 µL of EDTA whole blood
Дозволяє аналізувати зразок за більш ніж 50 різними гематологічним показниками
Proven, expert-class flow cytometry technology for leukocyte differential analysis across 6 subpopulations, including the analysis of immature granulocytes.
DESIGNED FOR VERSATILITY
Leukocyte counting and differential leukocyte count are performed using flow fluorescent cytometry.
Hemoglobin determination is performed using the high-precision SLS method.
The counting of erythrocytes and platelets is performed using the impedance method with hydrodynamic focusing.
Metrological traceability of measurement results for patient samples to reference methods is guaranteed by the manufacturer (ISO 17511).
Sysmex control materials are certified for all parameters specified for XN-L analyzers (XN-L CHECK)
Small sample volume (25 µL) with the capability for direct counting of immature granulocytes (IG) and antibody-producing cells (HFLC) in every analysis. Flagging and enumeration of nucleated red blood cells (NRBC).
The XN-L and XN series analyzers are the only instruments that provide the clinician with the absolute and relative counts of immature neutrophils (IG—the total count of metamyelocytes, promyelocytes, and myelocytes).

The XN-330 analyzer is an analyzer with manual sample loading in an open tube.
CHARACTERISTICS
Принцип вимірювання
Flow cytometry
Hydrodynamic focusing with the impedance method. SLS method for hemoglobin measurement
Parameters
RBC, HGB, HCT, MCHC, MCH, MCV, PLT, RDW-SD, RDW-CV, MPV, PDW, P-LCR, PCT, MicroR, MacroR, WBC, NEUT#/%, MONO#/%, EO# /%, BASO#/%, LYMPH#/%,IG#/%, NRBC#/%, HFLC#/%, NE-SSC, NE-SFL, NE-FSC
Пропускна здатність
CBC + DIFF – complete blood count with a 6-part white blood cell differential
60 samples/hour
Sample aspiration volume
Sample submission
Quality Control
25 µL of venous or capillary blood with EDTA
Manual sample loading in an open tube
Material (3 levels: XN-L CHECK L1, XN-L CHECK L2, XN-L CHECK L3) for quality control of all parameters determined in CBC+DIFF+RET mode
Fill volume: 3 ml per level
Reagents
CELLPACK DCL: 10/20 l
SULFOLYSER: 3 x 500 ml/5 l
LYSERCELL WDF: 2 l
FLUOROCELL WDF: 2 x 22 ml
CELLCLEAN: 50 ml
Dimensions
Dimensions: 450 x 460 x 510 mm
Weight: 35 kg